ESA615 Electrical Safety Analyzer
The ESA615 Electrical Safety Analyzer brings fast and simple automated testing in the form of a portable analyzer to healthcare technology professionals that perform electrical safety testing on medical equipment both in the field and in facilities. Whether it is simple testing or comprehensive analysis, the ESA615 can do it all. The ESA615 is an all-in-one solution with a safety analyzer and ECG simulator in a single electrical safety test instrument.
ESA615 Electrical Safety Analyzer
Test to any standard in minutes with the ESA615
Versatile to any global electrical safety standard of choice, the ESA615 can test to ANSI/AAMI ES-1, NFPA-99, IEC62353, VDE751, IEC60601-1 (2nd and 3rd editions) and AS/NZS 355 standards. The ESA615 is an all-in-one solution with a safety analyzer and ECG simulator in a single electrical safety test instrument. This multifaceted device performs all primary electrical safety tests, including line (mains) voltage, ground wire (protective earth) resistance, insulation resistance, device current, and patient lead leakage tests. It also offers ECG simulation and point-to-point voltage, leakage, and resistance tests.
Key Features
- Automated: Pre-set templates allow you to test to the global electrical safety standard of your choice at the push of a button. Let the unit guide you through the test steps quickly and accurately in three simple steps: 1.Select standard 2.Start test 3.Store results It is that easy. The automated test sequences are easily customized to suit individual testing requirements.
- Portable: The ESA615 is a small, lightweight analyzer complete with an integrated handle to make transportation from place to place simple for on-the-go field service professionals. It is designed for operation in tight spaces, and can easily fit on any cart. A light, protective carrying case makes it easy to store and transport safely.
- Easy to use: A push-button interface allows quick access to highly-comprehensive functions and features, and an intuitive interface guides the user through tests. The ESA615 features a large display for clear indication of available functions, set-up criteria, device under test receptacle conditions, and test results. Data can be entered quickly with a plug–and-play keyboard, barcode scanner and/or on-board data entry interface. Data archival is fast and simple with wireless communication or through a removable memory card with capacity for up to 100 automated test sequences and 1000 test results.
- Portable, ergonomic design with an integrated handle and tilt stand
- On-board automation with automated test sequences for easy compliance to key global electrical safety standards (ANSI/AAMI ES-1 (NFPA-99), IEC62353 (VDE751), IEC60601-1 2nd and 3rd editions, and AS/NZS 3551)
- User-friendly interface for streamlined testing
- ECG waveform tests and dual-lead measurements combine the functionality of a simulator and safety analyzer in a single test tool
- Five applied parts jacks and easy ECG snap connection; optional expander box for up to 12-lead ECG testing
- Easy data entry through barcode external keyboard or on-board keypad
- Wireless communication plus removable memory card for fast and convenient data storage and exchange
- 20 A at 120 V current capability
- Replaceable mains fuses keep your unit in the field and out of the repair shop
- Custom language selections include: English, French, German, Spanish, Italian and Portuguese
- Rigorously tested for rugged field applications with CE and CSA in addition to Fluke-quality- ruggedness for long-term reliability
- Two-year extended warranty (no-cost available after first-year calibration at any authorized Fluke Biomedical Service Center)
- Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide
Model Name | Part Number | Description |
---|---|---|
ESA615-USA | 4132046 | ESA615 US, 115V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) 15 – 20 A Adapter (2719-0154) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord US (LINE CORD) |
ESA615-01-FR | 4132054 | ESA615-01 France/Belgium, 230V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Europe (75026) |
ESA615-02-EUR | 4132093 | ESA615-02 Europe, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Europe (75026) |
ESA615-03-ISR | 4132101 | ESA615-03 Israel, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Israel (LINE CORD) |
ESA615-05-AUS | 4132112 | ESA615-05 Australia/China, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Australian (75025) |
ESA615-06-UK | 4132120 | ESA615-06 UK, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord UK (75024) |
ESA615-07-SWISS | 4132135 | ESA615-07 Switzerland, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Switzerland (FBC-ESA620-4420) |
ESA615-08-THAI | 4132147 | ESA615-08 Thailand, 230 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord US with C19 Plug (75033) |
ESA615-09-JAPAN | 4132158 | ESA615-09 Japan, 100 V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) 2195732 15-20A Adapter (2719-0154) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord USA (LINE CORD) |
ESA615-10-NA220V | 4132164 | ESA615-10 North America 220V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable USA Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT USA) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord US with C19 Plug (75033) Power Cord NA 220 V (75057) NA 220 V Adapter Kit (1800001) |
ESA615-11-BRAZ | 4161140 | ESA615-11 Brazil 230V Includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) 15 – 20 A Adapter (2719-0154) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord Brazilian (LINE CORD) |
ESA615-12-INDIA | 4161157 | ESA615-12-India, 250V includes: Operator’s Manual (multilingual CD) (CD-ROM) Getting Started Guide, hard copy, multilingual (Manual) Ansur Plug-in, CD with demo version (CD-ROM) Data Transfer Cable EUR Accessory Kit: Test Lead Set, TP1 Test Probe Set, AC285 Alligator Clip Set (ESA T/L KIT EUR) Null Post Adapter (ESA620-NPA) 5-to-5 Banana Jack to ECG (BJ2ECG) Adapter (ESA612-2016) Carrying Case (9530-0075) Power Cord India |
TA-ESA615-USA | 4162180 | ESA615 US, 115 V with test automation includes: ESA615-USA Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-FR | 4162198 | ESA615 France/Belgium, 230 V with test automation Includes: ESA615-01-FR Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-EUR | 4162211 | ESA615 Europe, 230 V with test automation Includes: ESA615-02-EUR Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-ISR | 4162227 | ESA615 Israel, 230 V with test automation Includes: ESA615-03-ISR Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-AUS | 4162230 | ESA615 Australia, 230 V with test automation Includes: ESA615-05-AUS Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-UK | 4162248 | ESA615 UK, 230 V with test automation Includes: ESA615-06-UK Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-SWISS | 4162253 | ESA615 SWISS, 230 V with test automation Includes: ESA615-07-SWISS Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-THAI | 4162275 | ESA615 THAI, 230V with test automation Includes: ESA615-08-SWISS Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-JAPAN | 4162282 | ESA615 JAPAN, 100 V with test automation Includes: ESA615-09-JAPAN Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-NA220V | 4162341 | ESA615 North America, 220 V with test automation Includes: ESA615-10-NA220V Ansur ESA615 Plug-In (ANSUR ESA615) |
TA-ESA615-12-INDIA | 4162365 | TA-ESA615-12-India, 250V with test automation includes: ESA615-12-India Ansur ESA615 Plug-In (ANSUR ESA615) |
Yes, test sequences can be created and modified per user needs
Many SD cards are compatible, but only the 512MB Industrial Grade SD card that ships with ESA615 has been tested and fully validated. LifeForce strongly encourages you to use the SD card supplied with the instrument. Additional SD cards are available at a minimal cost
No. The ESA615 does not have the ability to use a 4-wire method to measure earth conductor integrity. If this method is required, we recommend the ESA620.
Yes, except the limitations mentioned above (MAP source voltage 100 % of mains input and PE test current of 200 mA), the ESA615 tests according to IEC60601 2nd and 3rd editions.
Yes, it has a power supply that can accept both 120 V and 220 V; however the test receptacle is configured for one or the other.
Typically IP devices have no ground (no protective earth) and no power cord. That means the following cannot be done:
- No protective earth resistance
- No leakage to protective earth
- No single fault tests (there is no power cord---therefore no L1 or L2 wires to open)
That leaves enclosure leakage (touch current).
If the enclosure is plastic (no exposed metal test points) then use a 4 cm square of aluminum foil. This aluminum is taped to the surfaces of the medical device "touchable" by the patient or the clinician who is also touching the patient. Each such available surface must be tested and the maximum leakage reported. The test probe alligator clip is connected to the aluminum foil (which is why it has a certain thickness to prevent tearing), and the measurement is made and compared to the maximum leakage current allowed by the standard. IEC60601-1 and IEC 62353 each have slightly different requirements relative to power on and power off testing.
The ESA 615 is an automated electrical safety that has:
- Built-in automation that allows automatic testing to ANSI/AAMI ES-1 (NFPA-99), IEC62353 (VDE751), IEC60601-1 (2nd, 3rd edition), or AN/NZS 3551.
- Removable memory card with capacity for a minimum of hundreds of test sequences and thousands of test results
- Quick data entry options through plug €˜n' play keyboard, barcode scanner or on-board data entry interface
- Wireless data communication that makes remote operation and data archival fast and simple
- Custom language selections include English, French, German, Spanish, Italian and Portuguese
- An integrated handle that makes the unit easy to carry
The ESA615 has built-in test sequences for automatic electrical safety testing. These test sequences provide automatic pass/fail indication to NFPA-99 and other international standards. The unit guides you through the test steps quickly and accurately. Customizable test sequences cover ANSI/AAMI ES-1 (NFPA-99), IEC62353 (VDE751), IEC60601-1 (2nd, 3rd edition), or AN/NZS 3551. The ESA615 does not automate other test devices; however, it can be tethered to other test tools using Ansur automation software.
The ESA615 will attempt to load the last test sequence used on power up. If you have deleted it, for instance by reimaging your card, you will need to select a test sequences from the test library. Push F4 to say "OK" and then push F3 for "Test Library" and either select a test sequence or create a new one to resolve this error.
Be careful not to run automated tests from old firmware versions. You need to recreate any custom test sequences after updating to the latest firmware to receive all the benefits from the update.
- Avoid using Windows ReadyBoost. It will cause problems when the SD card is installed in an ESA615.
- Put the SD card into any USB memory card reader (not included).
- Save any test results (.DTA) files you wish to retain.
- Download the latest ESA615 Firmware Update from www.flukebiomedical.com under Support > Software Downloads.
- Verify your ESA615 firmware is up to date (look for it on power up or under Setup > Instrument Information). Update your firmware using the firmware update instructions, if needed.
- With the SD card installed on a Windows PC, follow the firmware update instructions to "Update SD card with new factory image". For firmware version 2.05, it is steps 13 to 25 in the firmware update instructions.
- ESA615 has the IEC60601 test load that can be selected by choosing this standard in the menu. However, it has the following limitations that do not allow for a fuller compliance with IEC60601:
- For Mains on Applied Part (MAP) testing, source voltage applied is only 100 % of Mains input voltage
- PE test current is 200 mA not 25 A
- The IEC60601-1:2005 standard now specifies earth leakage limit of 10,000 µA. Test equipment must be capable of measuring results throughout this range for equipment that is in the upper range.
- When using the 4-wire method, you are required to null the test lead resistance of your test lead; there is also a slight improvement in test accuracy when the 4-wire method is used. See specifications for detailed comparison.
- The IEC60601-1:2005 standard refers to patient and patient auxiliary leakage readings in both AC and DC parameters. If a safety analyzer only displays True RMS, you are not documenting all the results needed.
- The following test sequences are included with the ESA615:
- IEC60601-1 3rd Edition: Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device, Generic Device and System
- IEC62353: Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device
- NFPA-99 (Hospital): Patient Monitor, Defibrillator, Infusion Pump, Ultrasound Device and Generic Device
- ANSI/AAMI ES1: Patient Monitor, Defibrillator, Infusion Pump, Ultrasound
- This is a limitation of the file system on the SD card (FAT16/FAT32). You can avoid this behavior by naming your test with more than 8 characters.
1 ) Printer is not based on PCL5 or ZPLII
2 ) Every time the device and printer are powered off or the printer is connected to another device the printer needs to be reset.An example procedure to do so:
- ESA615 Reset the printer by:
- "Setup" -> F3 (printer settings) -> F1(Reset Printer)